NetSight
critical9.3

CVE-2013-2808

Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 comp…

Impact

Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 components, and Xper Information Management servers and workstations for Flex Cardio products before XperConnect 1.5.4.053 SP2 allows remote attackers to execute arbitrary code via a crafted HTTP request to the Connect broker on TCP port 6000.

Affected products

VendorProductAffected versions
PhilipsXper Information Management Physiomonitoring 5
All versions
PhilipsXperconnect
through 1.5.4.053
PhilipsXper Information Management Vascular Monitoring 5
All versions
PhilipsXper Flex Cardio
All versions

Weakness classification

References