NetSight
medium6.5

CVE-2017-12716

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programme…

Impact

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Affected products

VendorProductAffected versions
AbbottAccent Firmware
before f0b.0e.7e
AbbottAnthem Firmware
before f0b.0e.7e
AbbottAccent Mri Firmware
before f10.08.6c
AbbottAccent St Firmware
before f10.08.6c