CVE-2018-14781

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable…

Impact

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Affected products

VendorProductAffected versions
Medtronicdiabetes508 Minimed Insulin Pump Firmware
All versions
Medtronicdiabetes522 Paradigm Real-Time Firmware
All versions
Medtronicdiabetes722 Paradigm Real-Time Firmware
All versions
Medtronicdiabetes523 Paradigm Revel Firmware
All versions
Medtronicdiabetes723 Paradigm Revel Firmware
All versions
Medtronicdiabetes523k Paradigm Revel Firmware
All versions
Medtronicdiabetes723k Paradigm Revel Firmware
All versions
Medtronicdiabetes551 Minimed 530g Firmware
All versions
Medtronicdiabetes751 Minimed 530g Firmware
All versions

Weakness classification