NetSight
medium6.8

CVE-2018-18565

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000…

Impact

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.

Affected products

VendorProductAffected versions
RocheAccu-Chek Inform Ii Firmware
before 03.06.00
from 04.00.00 before 04.03.00
RocheCobas H 232 Firmware
before 03.01.03
from 04.00.00 before 04.00.04
RocheCoaguchek Pro Ii Firmware
before 04.03.00
RocheCoaguchek Xs Plus Firmware
before 03.01.06
RocheCoaguchek Xs Pro Firmware
before 03.01.06

Weakness classification

References