NetSight
medium6.4

CVE-2018-8870

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password.

Impact

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

Affected products

VendorProductAffected versions
Medtronic24950 Mycarelink Monitor Firmware
All versions
Medtronic24952 Mycarelink Monitor Firmware
All versions

Weakness classification

References