CVE-2019-10964
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, …
Impact
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected products
| Vendor | Product | Affected versions |
|---|---|---|
| Medtronic | Minimed 508 Firmware | All versions |
| Medtronic | Minimed Paradigm 511 Firmware | All versions |
| Medtronic | Minimed Paradigm 512 Firmware | All versions |
| Medtronic | Minimed Paradigm 712 Firmware | All versions |
| Medtronic | Minimed Paradigm 712e Firmware | All versions |
| Medtronic | Minimed Paradigm 515 Firmware | All versions |
| Medtronic | Minimed Paradigm 715 Firmware | All versions |
| Medtronic | Minimed Paradigm 522 Firmware | All versions |
| Medtronic | Minimed Paradigm 722 Firmware | All versions |
| Medtronic | Minimed Paradigm 522k Firmware | All versions |
| Medtronic | Minimed Paradigm 722k Firmware | All versions |
| Medtronic | Minimed Paradigm 523 Firmware | through 2.4a |
Weakness classification
References
- NVD record
- CVE.org record
- http://www.securityfocus.com/bid/108926
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed-508-paradigm.html
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-178-01