CVE-2019-6540
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer,…
Impact
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
Affected products
| Vendor | Product | Affected versions |
|---|---|---|
| Medtronic | Mycarelink Monitor 24950 Firmware | All versions |
| Medtronic | Mycarelink Monitor 24952 Firmware | All versions |
| Medtronic | Carelink Monitor 2490c Firmware | All versions |
| Medtronic | Carelink 2090 Firmware | All versions |
| Medtronic | Amplia Crt-D Firmware | All versions |
| Medtronic | Claria Crt-D Firmware | All versions |
| Medtronic | Compia Crt-D Firmware | All versions |
| Medtronic | Concerto Crt-D Firmware | All versions |
| Medtronic | Concerto Ii Crt-D Firmware | All versions |
| Medtronic | Consulta Crt-D Firmware | All versions |
| Medtronic | Evera Icd Firmware | All versions |
| Medtronic | Maximo Ii Crt-D Firmware | All versions |