medium4.3
CVE-2020-16220
In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be wel…
Impact
In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.
Affected products
| Vendor | Product | Affected versions |
|---|---|---|
| Philips | Patient Information Center Ix | Versions b.02, c.02, c.03 |
| Philips | Performancebridge Focal Point | Version a.01 |