NetSight
high7.6

CVE-2020-25166

An improper verification of the cryptographic signature of firmware updates of the B.

Impact

An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.

Affected products

VendorProductAffected versions
BbraunDatamodule Compactplus
Versions a10, a11
BbraunSpacecom
through l81

Weakness classification

References