critical9.3
CVE-2024-48971
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form.
Impact
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Affected packages
This record carries no package version data, so it cannot be matched against a dependency list automatically. Read the advisory and compare your version by hand.