NetSight
critical9.3

CVE-2024-48973

The debug port on the ventilator's serial interface is enabled by default.

Impact

The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected packages

This record carries no package version data, so it cannot be matched against a dependency list automatically. Read the advisory and compare your version by hand.