critical9.3
CVE-2024-9832
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password.
Impact
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected packages
This record carries no package version data, so it cannot be matched against a dependency list automatically. Read the advisory and compare your version by hand.