CVE-2018-8868

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, …

Auswirkung

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Ein Angreifer kann die Lücke mit physischem Zugriff auf das Gerät ausnutzen, mit einem normalen Benutzerkonto, ohne Zutun eines Nutzers. Dafür müssen bestimmte Bedingungen erfüllt sein, was einen Angriff erschwert. Bei Erfolg: hohes Risiko, dass Daten ausgelesen werden, geringes Risiko, dass Daten oder Systeme verändert werden, geringes Risiko für Ausfälle. Der Angriff kann über die verwundbare Komponente hinaus wirken.

Betroffene Produkte · Medtronic

ProduktBetroffene Versionen
24950 Mycarelink Monitor Firmware
Alle Versionen
24952 Mycarelink Monitor Firmware
Alle Versionen