Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmi…

Auswirkung

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Ein Angreifer kann die Lücke über das Netzwerk ausnutzen, ohne Anmeldung, ohne Zutun eines Nutzers. Bei Erfolg: hohes Risiko, dass Daten ausgelesen werden.

Betroffene Produkte · Baxter

ProduktBetroffene Versionen
Phoenix X36 Firmware
Versionen 3.36, 3.40