CVE-2023-25931

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requi…

Auswirkung

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Ein Angreifer kann die Lücke mit physischem Zugriff auf das Gerät ausnutzen, ohne Anmeldung, ohne Zutun eines Nutzers. Bei Erfolg: geringes Risiko, dass Daten ausgelesen werden, hohes Risiko, dass Daten oder Systeme verändert werden, hohes Risiko für Ausfälle.

Betroffene Produkte · Medtronic

ProduktBetroffene Versionen
Interstim X Clinician
Version a51300
Micro Clinician
Version a51200