In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be wel…

Auswirkung

In Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01, the product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling.

Ein Angreifer kann die Lücke aus demselben lokalen Netzwerk ausnutzen, ohne Anmeldung, ohne Zutun eines Nutzers. Bei Erfolg: geringes Risiko für Ausfälle.

Betroffene Produkte · Philips

ProduktBetroffene Versionen
Patient Information Center Ix
Versionen b.02, c.02, c.03
Performancebridge Focal Point
Version a.01